Testsystems
All articles

Know-how

What is a magnetic field camera?

A magnetic field camera is a sensor system that captures and visualizes the magnetic flux density across a flat surface. Industrial systems usually do this with a matrix of many small magnetic field sensors, typically Hall sensors, integrated together on one chip. The advantage of magnetic field cameras is that they capture the field distribution across a surface at high spatial resolution. And because the sensors are made on a common chip, their positions relative to each other are known very precisely.

What magnetic field cameras are used for

Magnetic field cameras are used above all to inspect permanent magnets, for example in incoming and outgoing inspection at companies that make or use magnets. The field image shows whether a magnet is evenly magnetized, in which direction it is magnetized, and whether the field shows irregularities that point to flaws such as cracks or chipped edges. On multipole magnets, it makes the pole pattern visible. In development, magnetic field cameras help to understand the field distribution of assemblies, for example to decide where a sensor should sit. Depending on the model, a magnetic field camera records only the field component perpendicular to the chip surface, or all three components.

A magnetic field camera plays to its strengths on small, flat parts whose measuring area fits on the chip. A single measurement then gives the complete field distribution, without moving either the part or the camera. On curved or large parts, on the other hand, the principle reaches its limits.

For our test benches, we decided against developing our own magnetic field camera, for several reasons.

Demanding calibration

First, calibrating a magnetic field camera takes a great deal of effort. A coil carrying a current (a Helmholtz coil, for example) generates a defined reference field. The readings of the camera’s sensors are then compared with those of a calibrated reference magnetometer. The difficult part is creating the same conditions for the reference measurement and the camera measurement, above all the same measuring position and the same distance from the field source. The camera housing makes this harder, because the exact position of the sensors inside it is hard to determine. Fields that change over time add a further problem: eddy currents form in the metal housing and damp and distort the field.

Slower sampling through multiplexing

Second, the measurement is comparatively slow, because the sensors are usually multiplexed. This means they are sampled one after another rather than at the same time. An array of individual sensors with their own measuring channels, by contrast, can be sampled synchronously. This matters especially for rotating parts: with multiplexing, the rotor keeps turning between the sampling of the first sensor and the last. Either the rotor stands still for each image, which costs time, or the image is not a snapshot. But even on flat surfaces, the low sampling rate of multiplexing slows the measurement down, especially when the camera has to be moved across the surface several times.

When a single sensor fails

Third, a failed sensor makes the measurement noisy: its signals appear as noise in the matrix. You can picture this much like dead pixels in the image of a digital camera. In that case, the complete chip has to be replaced. The sensors of a sensor array, by contrast, can be replaced individually.

The amplifier on the circuit board

Another challenge lies in the design. When laying out the circuit board, care must be taken that the amplifier affects the sensors as little as possible. The amplifier is needed to boost the weak signals of the sensitive sensors. If it sits too close to the sensors, however, it affects the measurement itself, for example through the magnetic fields of its supply currents, or through heat that makes the sensor readings drift. But it must not be too far away either, because long signal lines pick up interference.

Curved surfaces

Finally, measuring across a surface only plays to its strength on flat objects. On rotating parts such as a rotor, the surface is curved. The camera’s individual sensor rows are therefore at different distances from the rotor surface. Only the row with the smallest distance, the one parallel to the axis of rotation, gives comparable values. That largely cancels out the advantage of measuring across a surface.

Magnetic field camera above a rotor
Side view along the axis of rotation, schematic. The sensor rows run parallel to the axis, into the page.
Magnetic field camera above a rotor. Side view along the axis of rotation, schematic. The sensor rows run parallel to the axis, into the page. SNSNSNSN Magnetic field camera Rotor
  • Sensor row at the smallest distance: gives comparable values
  • Other sensor rows: farther from the rotor surface

How we measure instead

Our sensor array consists of up to 60 individual magnetic field sensors. It works as a line-scan magnetic field camera: as the rotor turns, the array builds up a field image of the entire surface line by line. On linear motor magnet tracks, the array scans the surface line by line and so constructs a signal of the whole area. For research and development, our 3D single sensor records all three magnetic field components at an angular resolution of 0.01°.

At a glance: magnetic field camera and sensor array

Magnetic field cameraSensor array
DesignA matrix of many Hall sensors on one common chipIndividual magnetic field sensors with their own measuring channels
Measuring areaA flat surfaceOne line; as the rotor turns, a field image of the entire surface builds up line by line
On a curved rotorThe sensor rows sit at different distances from the surface; only the row at the smallest distance gives comparable valuesThe sensor line runs parallel to the axis of rotation, every sensor at the same distance from the surface
SamplingUsually multiplexed, one sensor after another: a low sampling rate; if the rotor turns while the image is taken, it is not a snapshotSynchronous sampling
When a sensor failsThe faulty sensor makes the measurement noisy; the complete chip has to be replacedThe faulty sensor is replaced on its own
Best suited forSmall, flat parts, for example individual magnetsCurved, long and large parts: permanent magnet, externally excited (EESM) and axial flux rotors, and linear motor magnet tracks

Would you like to know more?

We measure the magnetic field of your rotors pole by pole, on one of our test benches or as a contract measurement.